Evaluation and routine testing in medical imaging departments - Part 3-5: Acceptance tests and and constancy tests - Imaging performance of computed tomography X-ray equipment (IEC 61223-3-5:2019)
This part of IEC 61223 applies to CT SCANNERS that conform to IEC 60601-2-44:2009,
IEC 60601-2-44:2009/AMD1:2012 and IEC 60601-2-44:2009/AMD2:2016.
IEC 60601-2-44 and this document
• defines the essential parameters which describe the performance of CT SCANNERS with
regard to image quality, RADIATION OUTPUT and PATIENT positioning; the list of parameters
to be tested can be found in 4.3,
• defines the methods of testing the essential parameters, and
• evaluates compliance with the tolerances of the parameters SPECIFIED by the
ACCOMPANYING DOCUMENTS.
The methods defined in IEC 60601-2-44 and this document rely on non-invasive
measurements, using appropriate test equipment, performed during or after installation.
Signed statements covering steps in the installation procedure can be used as part of the
ACCEPTANCE TEST report.
ΚΩΔΙΚΟΣ ΠΡΟΪΟΝΤΟΣ:
CYS EN IEC 61223-3-5:2019
This part of IEC 61223 applies to CT SCANNERS that conform to IEC 60601-2-44:2009,
IEC 60601-2-44:2009/AMD1:2012 and IEC 60601-2-44:2009/AMD2:2016.
IEC 60601-2-44 and this document
• defines the essential parameters which describe the performance of CT SCANNERS with
regard to image quality, RADIATION OUTPUT and PATIENT positioning; the list of parameters
to be tested can be found in 4.3,
• defines the methods of testing the essential parameters, and
• evaluates compliance with the tolerances of the parameters SPECIFIED by the
ACCOMPANYING DOCUMENTS.
The methods defined in IEC 60601-2-44 and this document rely on non-invasive
measurements, using appropriate test equipment, performed during or after installation.
Signed statements covering steps in the installation procedure can be used as part of the
ACCEPTANCE TEST report.