Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and essential performance of heating devices using blankets, pads and mattresses and intended for heating in medical use (IEC 60601-2-35:2020)
This part of 60601 International Standard applies to the BASIC SAFETY and ESSENTIAL
PERFORMANCE of HEATING DEVICES using BLANKETS, PADS or MATTRESSES in medical use, also
referred to as ME EQUIPMENT. HEATING DEVICES intended to prewarm a bed are included in the
scope of this document.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
If a clause or subclause is specifically intended to be applicable to a specifically defined type
of ME EQUIPMENT, as is the case with FORCED AIR DEVICES, then the title and content of that
clause or subclause will say so. If that is not the case, the clause or subclause applies both to
ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS
within the scope of this document are not covered by specific requirements in this document,
except in 7.2.13 and 8.4.1 of the general standard.
NOTE See also 4.2 of the general standard.
This document does not apply to:
-HEATING DEVICES intended for physiotherapy;
-INFANT RADIANT WARMERS; for information, see IEC 60601-2-21 [1]2;
- INFANT INCUBATORS; for information, see IEC 60601-2-19 [2];
-INFANT TRANSPORT INCUBATORS, for information, see IEC 60601-2-20 [3];
-cooling devices.
ΚΩΔΙΚΟΣ ΠΡΟΪΟΝΤΟΣ:
CYS EN IEC 60601-2-35:2021
This part of 60601 International Standard applies to the BASIC SAFETY and ESSENTIAL
PERFORMANCE of HEATING DEVICES using BLANKETS, PADS or MATTRESSES in medical use, also
referred to as ME EQUIPMENT. HEATING DEVICES intended to prewarm a bed are included in the
scope of this document.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
If a clause or subclause is specifically intended to be applicable to a specifically defined type
of ME EQUIPMENT, as is the case with FORCED AIR DEVICES, then the title and content of that
clause or subclause will say so. If that is not the case, the clause or subclause applies both to
ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS
within the scope of this document are not covered by specific requirements in this document,
except in 7.2.13 and 8.4.1 of the general standard.
NOTE See also 4.2 of the general standard.
This document does not apply to:
-HEATING DEVICES intended for physiotherapy;
-INFANT RADIANT WARMERS; for information, see IEC 60601-2-21 [1]2;
- INFANT INCUBATORS; for information, see IEC 60601-2-19 [2];
-INFANT TRANSPORT INCUBATORS, for information, see IEC 60601-2-20 [3];
-cooling devices.