Workplace atmospheres - Pumps for personal sampling of chemical and biological agents - Requirements and test methods
This International Standard specifies performance requirements for battery powered pumps used for
personal sampling of chemical and biological agents in workplace air. It also specifies test methods in order
to determine the performance characteristics of such pumps under prescribed laboratory conditions.
This International Standard is applicable to battery powered pumps having a nominal volume flow rate
above 10 ml ⋅ min−1, as used with combinations of sampler and collection substrate for sampling of
gases, vapours, dusts, fumes, mists and fibres.
This International Standard is primarily intended for flow-controlled pumps.
ΚΩΔΙΚΟΣ ΠΡΟΪΟΝΤΟΣ:
CYS EN ISO 13137:2013
This International Standard specifies performance requirements for battery powered pumps used for
personal sampling of chemical and biological agents in workplace air. It also specifies test methods in order
to determine the performance characteristics of such pumps under prescribed laboratory conditions.
This International Standard is applicable to battery powered pumps having a nominal volume flow rate
above 10 ml ⋅ min−1, as used with combinations of sampler and collection substrate for sampling of
gases, vapours, dusts, fumes, mists and fibres.
This International Standard is primarily intended for flow-controlled pumps.